5.1 Technical data

752 Pump Unit

19

5

Appendix

5.1

Technical data

Pump

Pump type

2-channel peristaltic pump with rotational speed

of

20/min (50 Hz)

24/min (60 Hz)

Pump capacity

typically 0.5

0.6 mL/min with

6.1826.050 pump tubing

Pressure

max. 4 bar (0.4 MPa)

Pumpable liquids

Clear liquids with no solid contents

Pump tubing material

PVC (Tygon

)

Mains connection

Mains voltage

115 V: 100...120 V

±

10 %

230 V: 220...240 V

±

10 %

Switchable with voltage selection insert in fuse

holder (see section 2.2.1)

Mains frequency

50...60 Hz

Power consumption

12 VA

Fuse

5 mm Ø, 20 mm length

100

120 V: 0.2 A (slow-blow)

220

240 V: 0.1 A (slow-blow)

Remote interface

Purpose

Remote control of pump (switching on/off)

Pin assignment

330

+5V

2

5

+

-

Pin

Safety specifications

Construction / Testing

According to IEC 1010 / EN 61010 / UL 3101-1,

protection class 1, degree of protection IP40

Safety directions

The Instructions for Use include information and

warnings to which the user must pay attention in

order to assure safe operation of the instrument.