Figure 3, Mounting the needles – Metrohm 885 Compact Oven SC User Manual

Page 22

Advertising
background image

3.4 Mounting the needles

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

885 Compact Oven SC

1

4

2

3

6.1805.060

6.2049.050

6.2816.080

6.2816.070

Figure 3

Mounting the needles

Mount the needles as follows:

1 Screw the 6.2049.050 needle holder into the distributor on the

guide head.

2 Screw the 6.2816.080 exhaust air needle onto the Luer connec-

tor of the needle holder.

3 Carefully guide the 6.2816.070 piercing needle into the opening

of the distributor from above and allow it to drop down.

Note

Take care to ensure that the white PTFE seal is positioned securely
on the needle.

4 Screw the 6.1805.060 FEP tubing by hand onto the opening of

the distributor.

5 Tightly screw the other end of the tubing to the gas outlet opening

(labeled with to sample) next to the tower, see figure.

Advertising