Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

885 Compact Oven SC

■■■■■■■■

21

5 Shift the height of the tip of the heating tube vertically in such a way

that the outlet opening of the tubing is immersed as deeply as possi-
ble. The tip must not however be permitted to get in the way of the
stirring bar in the KF titration cell. Afterwards give the septum stop-
per its final tightening.

3.11

Remote connections

The 885 Compact Oven SC can be used as a control device for a simple
automation system with a large variety of different instruments. Even
older Metrohm instruments can thus be integrated into an automated
analysis system.

3.11.1

Remote cable

The following connecting cable can be used with the 885 Compact Oven
SC:

6.2141.340 (9-pin/
25-pin)

For connections with a 756/831 coulometer, a 7xx Titrino or a Titrando
by means of a Remote Box 6.2148.010.

The cable transmits start and stop signals from the 885 Compact Oven SC
to the connected titrator and status signals (Cond OK, EOD) from the titra-
tor to the 885 Compact Oven SC.

3.11.2

Example systems

The following illustrations show typical automation systems with different
instrument combinations.

885 — 756/831 Coulometer/7xx Titrino
The standard combination for water determinations with a 7xx Coulome-
ter or a 7xx KF Titrino.