2 reports, Reports – Metrohm 885 Compact Oven SC User Manual
Page 63
■■■■■■■■■■■■■■■■■■■■■■
7 Parameters
885 Compact Oven SC
■■■■■■■■
53
7.2
Reports
Menu
▶ Parameters ▶ Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains information concerning oven temperature and
gas flow during the determination.
Selection
on | off
Default value
off
Parameters
All of the parameters of the current method are printed out with the para-
meter report.
Selection
on | off
Default value
off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 47).
The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt.
Selection
on | off
Default value
off