5 assembling the drying flasks, Assembling the drying flasks, Figure 4 – Metrohm 885 Compact Oven SC User Manual

Page 23: Connecting the tubing to the gas outlet, Figure 5, Preparing the drying flasks

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

885 Compact Oven SC

■■■■■■■■

13

Figure 4

Connecting the tubing to the gas outlet

3.5

Assembling the drying flasks

Two drying flasks with desiccant are integrated into the gas flow in order
to dry the gas that is conveyed. Dust (e.g. from the desiccant) must be
prevented from finding its way into the sample vessel.

1

2

3

6.1608.050

6.1821.040

6.1602.145

6.2811.000

Figure 5

Preparing the drying flasks

Prepare both drying flasks as follows:

1 Fill both 6.1608.050 drying flasks with 6.2811.000 molecular

sieve.

2 Screw one 6.1821.040 filter tube into each of the 6.1602.145

drying flask covers from below. Tighten the filter tubes well by
hand.

3 Screw the two drying flask covers with the filter tubes onto the dry-

ing flasks. Tighten the covers well by hand.

Advertising