10 Appendix
885 Compact Oven SC
■■■■■■■■
63
Proceed as follows for system initialization:
1 Switch off the instrument
■
Keep the red [STOP] key pressed down for at least 3 s.
A progress bar is displayed. If the key is released during this time,
then the instrument will not be switched off.
2 Switch on the instrument
■
Keep the red [STOP] key pressed down for approx. 10 s.
The dialog for confirmation of the initialization is displayed for 8 s.
The initialization must be confirmed during this time.
3 Confirm the initialization
Note
If the query is not confirmed within 8 s, then the procedure will be
interrupted.
■
Press [BACK] twice.
Initialization is started. The process takes approximately 80 s. The
instrument will be automatically restarted after successful initializa-
tion.