2 remote interface, 1 pin assignment of the remote interface, Pin assignment of the remote interface – Metrohm 885 Compact Oven SC User Manual
Page 71: Figure 22, Pin assignment of remote socket and plug
Advertising
■■■■■■■■■■■■■■■■■■■■■■
10 Appendix
885 Compact Oven SC
■■■■■■■■
61
10.2
Remote interface
10.2.1
Pin assignment of the remote interface
1
5
6
9
5
1
9
6
Figure 22
Pin assignment of remote socket and plug
The above figure of the pin assignment applies for all Metrohm instru-
ments with 9-pin D-Sub remote connectors.
Table 2
Inputs and outputs of the remote interface
Pin No.
Assignment
Function
1
Output 0
Sample Ready
2
Output 1
Temperature ok
3
Output 2
Flow on
4
Output 3
5
Output 4
Error
6
0 volt (GND)
7
+5 volts
8
Input 0
Continue
9
Input 1
Cond ok
Inputs
+5 V
approx. 5 k
Ω Pull-up
tp
t
p
> 100 ms
active = low, inactive = high