7.1 Automation

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

885 Compact Oven SC

Only when 'Gas type' = 'other'.
Range

0.001 ... 2.000 (Increment: 0.001)

Default value

1.000

Table 1

Measuring factors of different gases

Gas type

Measuring
factor

Argon

0.950

Helium

0.300

Oxygen

1.000

Air

1.000

Nitrogen

1.000

End of series

Behavior of the instrument at the end of the sample series.

Selection

Stop | Conditioning

Default value

Stop

Stop
After the last sample, the entire system is conditioned one more time.
Afterwards, the needle is moved out of the conditioning vessel. The
gas flow and the oven heating are switched off. The titrator is stopped.
Conditioning
After the last sample, the entire system is conditioned one more time.
Afterwards, the needle is moved out of the conditioning vessel. The
gas flow and the oven heating are switched off. The titrator is kept in
conditioned status.

Conditioning timeout

This setting determines the maximum waiting time for conditioning. If no
stable conditioning signal is obtained from the titrator during this waiting
time for a total duration of 60 s, then an error message will be displayed,
after which the sample series will be stopped.

Range

2 ... 120 min (Increment: 1)

Default value

30 min