2 cleaning the spm, Figure 53, Spm – components – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual

Page 109

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

850 Professional IC Anion – MSM-HC – MCS – Prep 3

■■■■■■■■

99

3 Connecting the SPM to the IC system

Reconnect the SPM to the IC system. If the capacity problems
remain, the SPM rotor A must be replaced (see Chapter 5.10.3.3,
page 101)
.

5.10.3.2

Cleaning the SPM

It may be necessary to clean the SPM in the following cases:

Increased backpressure on the connection tubings of the SPM.

Blockage of the SPM which cannot be eliminated (solutions can no lon-
ger be pumped through the SPM).

Jamming of the SPM which cannot be eliminated (SPM can no longer
be switched over).

1

2

3

4

5

Figure 53

SPM – Components

1

Union nut

2

SPM connecting piece (6.2835.010)

3

SPM rotor A (6.2835.000)
Cation exchanger.

4

SPM housing

5

Slot in the SPM housing

SPM cleaning

Clean the SPM as follows:

1 Disconnecting the SPM from the system

Switch off the instrument.

Disconnect capillary connections of the SPM from the system.

2 Dismounting SPM

Unscrew the union nut (53-1) from the SPM housing (53-4).

Advertising