6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

120

■■■■■■■■

850 Professional IC Anion – MSM-HC – MCS – Prep 3

Problem

Cause

Remedy

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.16.4, page 114)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 75)
.

Eluent – Gas bubbles in the
eluent.

Check connections of the eluent degasser (see
Chapter 3.9, page 36)
.

High pressure pump –
defective.

Request Metrohm Service.

Vacuum is not being
built

Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.

Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module:

Connect an FEP tubing between the IC
instrument and the Extension Module and
tighten both clamping screws until they
seal.

SPM – No (or insuffi-
cient) pumping of
regeneration or rins-
ing solution

Leak in the system.

Check connections.

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure .

Peristaltic pump – filter
blocked (see Chapter 3.15,
page 52).

Replace the filter .

SPM – backpressure too
high.

Clean the SPM (see Chapter 5.10.3.2, page
99)
or replace parts (see Chapter 5.10.3.3,
page 101)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing .