Figure 25, Cation exchange with the spm, Figure 26 – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual
Page 59: Injection and sequential suppression
Advertising
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
850 Professional IC Anion – MSM-HC – MCS – Prep 3
■■■■■■■■
49
1
A
B
D
C
Figure 25
Cation exchange with the SPM
1
Sample
A
SPM
B
Injection valve
C
Separation column
D
MSM-HC
2 Injection and sequential suppression
■
The injection valve is switched to the position Inject.
■
With the eluent flow (26-2) the sample prepared is rinsed onto
the separation column (26-C).
■
The sequential suppression with the MSM-HC (26-D) and the
MCS follows after the separation.
A
B
D
C
2
Figure 26
Injection and sequential suppression
2
Sample in the eluent flow
A
SPM
B
Injection valve
C
Separation column
D
MSM-HC
Neutralization
A typical use of the SPM is the neutralization of an alkaline sample for
anion determination with chemical suppression.