5.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

90

■■■■■■■■

850 Professional IC – Cation – Prep 2

Problem

Cause

Remedy

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

Chromatograms
have poor resolution

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 4.13.4, page 87)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 60)
.

Conductivity detec-
tor is not recognized
in the software

No connection.

Check the cable connection (32-

1

).

Turn the instrument off and on again (after
15 seconds).

Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)

Connections – dead vol-
ume in the system.

Check the connections (see Chapter 2.5, page
17)
(use PEEK capillaries with an internal diam-
eter of 0.25 mm between the injection valve
and detector).

Guard column – diminished
performance.

Replace guard column (see Chapter 2.19,
page 57)
.

Separation column – dead
volume at column head.

Install separation column in reverse flow
direction (if permitted by the information
sheet).

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 60)
.

Precision problems -
significant scatter-
ing of the measured
values

Sample – Gas bubbles in
the sample.

Use sample degasser (see Chapter 2.13, page
43)
.

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 2.15.1, page 49)
.

Sample – Rinsing volume
too low.

Prolong the rinsing time (see Chapter 4.8,
page 80)
.

Injection valve – defective.

Request Metrohm Service.

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 4.13.4, page 87)
.