6 Technical specifications

850 Professional IC – Cation – Prep 2

■■■■■■■■

95

6.10

Injection valve

Actuator time

typ.100 ms

Max. operating
pressure

35 MPa (350 bar)

Material

PEEK

6.11

Column thermostat

Type

Peltier technique thermostat for two intelligent separation columns

Adjustable tem-
perature range

0…+ 80 °C, in increments of 0.1 °C

Heating

Ambient temperature +50 °C

Cooling

Ambient temperature –20 °C

Temperature
reproducibility

± 0.2 °C

Stability

< 0.05 °C

Heating up time

< 30 minutes from 20 to 50 °C

Cooling time

< 40 minutes from 50 to 20 °C

6.12

Conductivity measuring system

Type

Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)

Intelligent detector with 6 reference chromatograms

Measuring range

0…15000 µS/cm single range

Noise

< 0.1 nS at 1 µS/cm

Drift

< 0.2 nS/cm per hour

Measuring rate

10 measurements per second for optimum results without filtering

Resolution

0.0047 nS/cm

Baseline

Noise < 3 nS/cm typical without suppression

Conductivity
detector

Cell volume

0.8 µL

Cell constant

Individual calibration data saved in the detector

adjustable range: 13.0…21.0 /cm