5 Troubleshooting

850 Professional IC – Cation – Prep 2

■■■■■■■■

91

Problem

Cause

Remedy

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 60)
.

Eluent – Gas bubbles in the
eluent.

Check connections of the eluent degasser (see
Chapter 2.9, page 34)
.

High pressure pump –
defective.

Request Metrohm Service.

Vacuum is not being
built

Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.

Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module:

Connect an FEP tubing between the IC
instrument and the Extension Module and
tighten both clamping screws until they
seal.