2 Installation

850 Professional IC – Cation – Prep 2

■■■■■■■■

13

3

High pressure pump
See Chapter 2.10.

4

Purge valve
See Chapter 2.10.1.

5

Inline filter
See Chapter 2.11.

6

Pulsation damper
See Chapter 2.12.

7

Injection valve
See Chapter 2.15.

8

Separation column
See Chapter 2.20.
When a guard column (see Chapter 2.19,
page 57)
is used, it is installed between
injection valve and separation column.

9

Detector
See Chapter 2.17.

10 Eluent output

Connection to the waste container.

11 Sample input

12 Injection valve

For sample preparation. See Chapter 2.15.

13 Sample output

14 Sample degasser

See Chapter 2.13. Use optional.

15 Transfer solution inlet

Connection to transfer solution container.

16 Peristaltic pump

Pumps the transfer solution. See Chapter
2.14
.

17 Preconcentration column

For sample preconcentration.

18 Transfer solution outlet

Connection to the waste container.

Standard addition with preconcentration
It is often the case that there are no certified standards available for com-
plicated sample matrices. "Standard addition with preconcentration" tech-
nology can be applied in order to prevent adulterations of the analysis
results caused by the matrix.

Procedure:

1 Sample measurement

The sample is rinsed for a defined time period onto the preconcen-
tration column (3-13) at the injection valve (3-7) and retained there.
Then the injection valve (3-7) is switched to <Inject> and data
recording is started. The eluent carries along the sample retained on
the preconcentration column (3-13) to the separation column.

2 1. Standard addition

Note

The sample path should be rinsed between measurements.