background image

6.13 Mains connection

■■■■■■■■■■■■■■■■■■■■■■

96

■■■■■■■■

850 Professional IC – Cation – Prep 2

Electrodes

Ring-shaped electrodes made from stainless steel

Materials in
contact with
eluent

Chemically inert PCTFE

Maximum oper-
ating pressure

5.0 MPa (50 bar)

Cell tempera-
ture

20…50 °C in increments of 5 °C

Temperature
stability

< 0.001 °C

Temperature
compensation

0…5 %/K adjustable, default 2.3 %/K

Heating up time

< 30 minutes (40 °C)

6.13

Mains connection

Required voltage

100…240 V ± 10 % (auto-sensing)

Required fre-
quency

50…60 Hz ± 3 (auto-sensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

internal fuse 3.15 A

6.14

Interfaces

USB

Input

1 USB upstream, type B (for connection to the PC)

Output

2 USB downstream, type A

MSB

2 MSB Mini Din 8-pin (female) (for Dosino, stirrer, remote lines, ...)

Caution

When connecting an instrument to the MSB connector you must
switch off the 850 Professional IC .

Detector

2 DSUB-15-pin high density (female)

Column recogni-
tion

3 (including 2 in the column thermostat (see Chapter 2.16, page 51))

Leak sensor

1 jack plug