3.1 Inserting the working electrode

■■■■■■■■■■■■■■■■■■■■■■

4

■■■■■■■■

IC Equipment Thin-Layer cell

3 Installation

Caution

Never switch on the measuring cell..

... if it is not simultaneously being rinsed by a conductive eluent; or

... if the electrode cables are not completely connected.

3.1

Inserting the working electrode

The measuring cell is supplied without electrodes. The working electrode
must be ordered separately.

Insert the working electrode as follows:

Inserting the working electrode

1 Unscrew the pressure screw on the base part of the cell and remove

it.

2 Insert the working electrode in the opening. The working electrode is

formed in such a way that it can be inserted into the measuring cell
in only one position.

3 Slide the pressure screw over the working electrode and screw it

tightly.