Metrohm Thin-Layer cell User Manual

Page 24

Advertising
background image

5.2 Maintenance

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

IC Equipment Thin-Layer cell

The working electrode has been removed from the measuring cell.

1

Note

When cleaning the auxiliary electrode, the edges of the spacer
may get damaged.

The spacer must then be replaced.

Moisten a cotton swab with 2 mol/L of nitric acid.

Use this to carefully wipe the auxiliary electrode.
The edges of the spacer may be damaged when wiping the auxili-
ary electrode.
If the auxiliary electrode is tenaciously stained, we recommend
that the spacer be removed prior to cleaning the auxiliary elec-
trode (see Chapter 5.2.3, page 15).

2 Rinse off the measuring cell (except for the working electrode and

the reference electrode) under running water and dry with a lint-free
cloth.

Note

When you rinse off the measuring part, take care to ensure that
the chip in the cell holder does not get wet.

Cleaning the working electrode

1 Clean the working electrode in accordance with the directions con-

tained in the leaflet for the working electrode or polish it if necessary.

2 Reinsert the working electrode (see Chapter 3.1, page 4).