Metrohm Thin-Layer cell User Manual

Page 15

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

IC Equipment Thin-Layer cell

■■■■■■■■

9

3 Connect the cell output

Use a pressure screw (6.2744.014) to fasten a 1–1.5 m long piece of
the PEEK capillary (6.1831.010) to the Out connector on the measur-
ing cell.

Note

This capillary must be tested with the IC instrument the first time
the detector is put into operation, see Chapter "Start-up" in the
manual for the detector.

If the preheating capillary is not used, proceed as follows:

Connecting capillaries to the measuring cell

1 Connect the cell input

Use a pressure screw (6.2744.014) to fasten the detector input capil-
lary to the In connector on the measuring cell.

2 Connect the cell output

Use a pressure screw (6.2744.014) to fasten a 1–1.5 m long piece of
the PEEK capillary (6.1831.010) to the Out connector on the measur-
ing cell.

Note

This capillary must be tested with the IC instrument the first time
the detector is put into operation, see Chapter "Start-up" in the
manual for the detector.