5.2 Maintenance

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

IC Equipment Thin-Layer cell

1 Unscrew the pressure screw (1-4) on the base part of the cell and

remove it.

2 Take out the working electrode.

3 Insert the new working electrode. The working electrode is formed in

such a way that it can be inserted into the measuring cell in only one
position.

4 Slide the pressure screw over the working electrode and screw it

tightly.

5.2.2

Replacing the reference electrode

Note

These instructions apply analogously to the Ag/AgCl reference elec-
trode.

Inserting the reference electrode

Prerequisites:

The measuring cell is switched off.

The measuring cell has been removed from the detector.

The electrode cables are unplugged.

No tool is required for replacing the reference electrode.

1 Slide the supplied sealing ring over the new reference electrode.