PROCEDURE

Use universal sample holder.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TEST from menu.
3. Scroll to and select ALL TESTS (or another sequence containing

29 Copper BCA-LR) from TESTING MENU.

4. Scroll to and select 29 Copper BCA-LR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with

sample.

6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from Spectro and add one *Copper Tablet (T-3808). Cap and

shake vigorously until tablet dissolves. Solution will turn purple if copper is

present. Wait 2 minutes.

8. At end of 2 minute waiting period, mix, insert tube into chamber, close lid and

select SCAN SAMPLE. Record result.

9. Press

OFF button to turn spectrophotometer off or press EXIT button to exit to

a previous menu or make another menu selection.

COPPER, BCA

SMART Spectro Test Procedures 2.11