APPLICATION:

Swimming pool water.

RANGE:

0–150 Cyanuric Acid

MDL:

5 ppm

METHOD:

A buffered solution of melamine forms a precipitate with

cyanuric acid in proportion to the amount of cyanuric

acid present. The amount of particles in suspension is

measured turbidimetrically.

SAMPLE HANDLING

& PRESERVATION:

Cyanuric acid samples should be analyzed as soon as

possible after collection. Deterioration of the sample

can be minimized by keeping samples in the dark or

refrigerated until analysis can be performed.

INTERFERENCES:

No known interference from compounds normally found

in pool water. Temperature of the sample should be

maintained between 70°F and 80°F for best results.

Check for stray light interference (see page 16).