SMART Spectro Operator’s Manual 2.11

41

6. Place a blank in the sample

chamber. Press

*ENTER to SCAN

BLANK. (STD. 1 is always a blank.)

The display will indicate that the

blank is being scanned. The

display will indicate BLANKED. A

cursor will appear on the character

selection line.

ENTER STD. 1

*

SCAN BLANK
STD.1 =

!

0 1 2 3 4 5 6 7 8 9 .

7. Place the reacted standard of

the lowest concentration in the

chamber. Press

or

to enter

the concentration of the standard

by selecting the characters with

the cursor.

ENTER STD. 1

*

SCAN BLANK
STD.1 =

!

0

1 2 3 4 5 6 7 8 9

8. After the concentration has been

entered, scroll to !. Press

*ENTER.

The standard will be scanned

and the absorbance will be

displayed.

ENTER STD. 1

*

BLANKED
STD.1 = 1.0
A=0.0016A

9. Press

EXIT to return to the New

Calibration menu.

NEW CALIBRATION

*

ENTER STD. 1
ENTER STD. 2
ENTER STD. 3
ENTER STD. 4

10. Press

ENTER and use the same

procedure to scan the second

data point. Repeat the procedure

until the data from all of the

standards has been entered into

the calibration.

Note: Input the number of data

points previously selected.

For example, when inputting 6

standards ignore the request for

ENTER STD. 7.

NEW CALIBRATION
ENTER STD. 4
ENTER STD. 5
ENTER STD. 6

*

ENTER STD. 7