SMART Spectro Operator’s Manual 2.11
45
4. Press
*ENTER to select SELECT
WL
SELECT WL
WL=560 NM
*
5. Press or
to scroll to desired
wavelength.
*
SELECT WL
WL=600 NM
6. Press
*ENTER to select the
wavelength. The meter is ready to
scan.
%T/ABS 600 NM
SELECT WL
*
SCAN BLANK
SCAN SAMPLE
END OF LIST
7. Insert the blank into the chamber.
Press
*ENTER to scan the blank.
Note: For most %T/ABS tests, a
clear colorless water blank should
be used.
%T/ABS 600 NM
SELECT WL
SCAN BLANK
*
SCAN SAMPLE
END OF LIST
8. Insert a reacted sample into the
chamber. Press
*ENTER to scan
the sample.
%T/ABS 600 NM
T=90.7%T
A=0.0424A
PRINT PRESS ENTER
At this point, it is possible to scan another sample, scan another blank, or select
another wavelength. To print the results to a connected computer or serial
printer, press
*ENTER and return to previous menu. Press EXIT to escape to
previous menu.
If no more samples are to be scanned, press
EXIT to return to the MAIN MENU
or press
OFF to turn off the spectrophotometer.