APPLICATION:

Drinking, surface, and saline waters; domestic and

industrial wastes.

RANGE:

0.00–1.00 ppm Sulfi de

MDL:

0.02 ppm

METHOD:

Under suitable conditions the sulfi de ion reacts

with p-aminodimethylaniline and ferric chloride to

produce methylene blue in proportion to the sulfi de

concentration. Ammonium phosphate is added to

remove the color due to the ferric iron.

SAMPLE HANDLING

& PRESERVATION:

Samples must be taken with a minimum of aeration

since sulfi de is volatilized by aeration and any oxygen

which is taken up will destroy sulfi des by chemical

action. Samples that are used for total sulfi de

concentrations may be preserved by adding 2M

zinc acetate solution at a dosage of 2 mL per liter of

sample. This precipitates sulfi de as inert zinc sulfi de.

Determination of dissolved sulfi des in samples not

preserved with zinc acetate must be started within 3

minutes of sampling.

INTERFERENCES:

Strong reducing agents such as sulfi te, thiosulfate,

and hydrosulfi te prevent the formation of the color or

diminish its intensity. High concentrations of sulfi de will

inhibit the reaction, but dilution of the sample prior to

analysis eliminates this problem.