LaMotte SMART Spectro Spectrophotometer User Manual
Page 175
Advertising
APPLICATION:
Drinking, surface, and saline waters; domestic and
industrial wastes.
RANGE:
0.00–10.00 ppm
MDL:
0.07 ppm
METHOD:
Ferric iron is reduced to ferrous iron and subsequently
forms a colored complex with for a quantitative measure
of total iron.
SAMPLE HANDLING
& PRESERVATION:
The sample container should be cleaned with acid and
rinsed with deionized water. Addition of acid to adjust
the sample. The pH 2-3 will prevent depositation of iron
on the container walls. Samples should be analyzed as
soon as possible.
INTERFERENCES:
Strong oxidizing agents interfere, as well as copper and
cobalt in excess of 5.0 ppm.
IRON, UDV
SMART Spectro Test Procedures 2.11
Test P
rocedures