LaMotte SMART Spectro Spectrophotometer User Manual
Page 42
42
SMART Spectro Operator’s Manual 2.11
11. Press
to scroll to
CALCULATE.
NEW CALIBRATION
ENTER STD. 6
ENTER STD. 7
ENTER STD. 8
*
CALCULATE
12. Press
*ENTER to select
CALCULATE.
SELECT DEGREES
1 DEGREE THRU.0
1 DEGREE
2 DEGREE
*
3 DEGREE
13. Press
*ENTER to select the
desired curve fi t.
SELECT DEGREES
1 DEGREE THRU.0
*
1 DEGREE
2 DEGREE
3 DEGREE
Note:
1 DEGREE THRU.0 calculates the best straight line fi t through the data
points and intersects with the origin at 0 ppm, 0 absorbance. This is a classical
Beers Law calibration. If a one point calibration has been performed, the
1
DEGREE THRU.0 curve fi t must be chosen. 1 DEGREE calculates the best
straight line fi t but without forcing the line through the origin. The minimum
number of standards needed for a calibration is one for
1 DEGREE THRU.0 and
two for
1 DEGREE.
NOTE: 2 DEGREE and 3 DEGREE calculations are not available yet.
14. Press
*ENTER to select the
desired curve fi t type. The display
will show the constraints for the
best fi t line.
KO=0.0000E +00
K1=0.3068E +0A
K2=0.0000E +00
K3=0.0000E+0A
Press EXIT return