LaMotte TC-3000 Tri-Meter User Manual

Page 15

Advertising
background image

5. Scroll down and then press

*IOK

to select a chlorine reagent system.

The options are: Tablet or Liquid

.

C a l i b r a t i o n
* T a b l e t

L i q u i d

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

6. Press

… to exit to a previous menu or

make another menu selection or
press

OFF to turn the meter off.

AVERAGING

The averaging option is available for color and turbidity testing. It allows the user to
average multiple readings. This option will improve the accuracy of samples with
readings that may tend to drift with time. When the two, five or ten reading options
have been selected, the meter will show a running average of the readings that have
been taken until the final average is displayed. The default setting is disabled. To
change the setting:

1. Press

ON to turn the meter on.

1 . 3

2. Scroll down and then press

*IOK

to select Options.

M a i n M e n u

M e a s u r e
D a t a L o g g i n g

* O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

3. Press

*IOK to select Averaging.

O p t i o n s
* A v e r a g i n g

T u r b i d i t y
C h l o r i n e
D a t e / T i m e

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

OPTIONS & SETUP • AVERAGING

15

Advertising