LaMotte TC-3000 Tri-Meter User Manual

Page 29

Advertising
background image

4. Rinse a clean tube (0290) three

times with the blank.

Below 1 NTU – The meter should be
blanked with a 0 NTU Primary
Standard or prepared turbidity-free
(<0.1 NTU) water. For the most
accurate results, use the same tube
for the blank and the sample.

5. Fill the tube to the fill line with the

blank. Pour the blank down the
inside of the tube to avoid creating
bubbles. Cap the tube.

6. Dry the tube with a lint-free cloth. Put

on a dry positioning ring. Cap the
tube. Wipe the tube thoroughly
again with a lint-free cloth.

7. Open the meter lid. Insert the tube

into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter. Close
the lid.

8. Press

*IOK to select Scan Blank.

Remove the tube.

Note: For the best accuracy,

especially at low turbidity, see Tip 17
on page 33.

T u r b i d i t y

( F )

* S c a n B l a n k
1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

ANALYSIS & CALIBRATION • TURBIDITY

29