LaMotte TC-3000 Tri-Meter User Manual

Page 58

Advertising
background image

13. When the value on the display

matches the concentration of the
color standard, press

*IOK to select

Set

.

C a l i b r a t e

0 3 0 . 0

* S e t

€

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

Or press the

ˆ and press *IOK

to return the meter to the default
setting.

C a l i b r a t e

0 3 0 . 0

* D e f a u l t

€

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

14. Press

*IOK to proceed to Turbidity

analysis. Press

OFF to turn the meter

off or press

… to exit to a previous

menu or make another menu
selection.

Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tube positioning ring should be used for the blank and the reacted sample.

Testing Tips

1.

Always use a clean test tube.

2.

Use the averaging option for low level measurements of color.

3.

Turbidity will interfere with the color test. Sample may be filtered before testing but
results will be as true color. (See

What is Color?

, page 22)

4. When testing at low concentrations use the same tube for the blank and the

sample.

5. Always use the positioning ring. Always insert tube into the meter chamber with the

same amount of pressure and to the same depth.

6. Occasionally clean the chamber with a damp lint-free wipe, followed by an

alcohol dampened wipe. A clean chamber and tubes are essential for reliable
results.

58

ANALYSIS & CALIBRATION • COLOR