LaMotte TC-3000 Tri-Meter User Manual

Page 37

Advertising
background image

2. Press

*IOK to select Measure.

M a i n M e n u
* M e a s u r e

D a t a L o g g i n g
O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

3. Scroll down and then press

*IOK

to select Chlorine.

M e a s u r e

T u r b i d i t y
C o l o r

* C h l o r i n e

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

4. Scroll down and press

*IOK to

select Test Total Chlorine.

C h l o r i n e

T e s t F r e e

* T e s t T o t a l

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

5. Rinse a clean tube (0290) with the

sample water. Fill the tube to the
10 mL line with the sample water.
Dry the tube with a lint-free cloth.
Put on a dry positioning ring. Cap
the tube.

6. Open the meter lid. Insert the tube

into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter. Close
the lid and press

*IOK to select

Scan Blank

.

T o t a l C h l o r i n e

( T )

* S c a n B l a n k

‚

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

7. Remove the tube from the meter.

Pour off all but a sufficient amount of
sample water to cover a tablet. Add
one *Chlorine DPD #1 Instrument
Grade Tablet (6903). Crush tablet
with a tablet crusher (0175), then
add sample water until the tube is
filled to the 10 mL line. Cap tube
and shake until tablet has dissolved.

ANALYSIS & CALIBRATION • CHLORINE

37