Turbidity Standards

Only use AMCO or formazin standards with the TC-3000. StablCal

â standards

below 50 NTU should not be used to calibrate the TC-3000. The diluent used in
StablCal

â

standards has a different refractive index than traditional formazin standards

and will affect the results. The concentration of the calibration standard should be
similar to the expected concentration of samples that will be tested. The following
standards are available from LaMotte Company:

1480

0 NTU Standard, 60 mL (EPA and ISO)

1484

1 NTU Standard, 60 mL (EPA)

1481

1 NTU Standard, 60 mL (ISO)

1485

10 NTU Standard, 60 mL (EPA)

1482

10 NTU Standard, 60 mL (ISO)

1486

100 NTU Standard, 60 mL (EPA)

1483

100 NTU Standard, 60 mL (ISO)

Calibration Procedure

1. Press

ON to turn the meter on.

1 . 3

2. Press

*IOK to select Measure.

M a i n M e n u
* M e a s u r e

D a t a L o g g i n g
O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

3. Press

*IOK to select Turbidity.

M e a s u r e
* T u r b i d i t y

C o l o r
C h l o r i n e

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

28

ANALYSIS & CALIBRATION • TURBIDITY