LaMotte TC-3000 Tri-Meter User Manual
Page 34
Advertising
3. Scroll down and then press
*IOK
to select Chlorine.
M e a s u r e
T u r b i d i t y
C o l o r
* C h l o r i n e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
4. Press
*IOK to select Test Free
Chlorine.
C h l o r i n e
( T )
* T e s t F r e e
T e s t T o t a l
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
5. Rinse a clean tube (0290) with the
sample water.
6. Fill the tube to the 10 mL line with
the sample water. Dry the tube with
a lint-free cloth. Put on a dry
positioning ring. Cap the tube.
7. Open the meter lid. Insert the tube
into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter.
34
ANALYSIS & CALIBRATION • CHLORINE