5.

Connect the mains cable to the M3 terminal in the termination
area of the converter.

6.

Allow a little bit of slack for all cables (i.e., sensor, power, and
digital) and tighten all the cable glands on the outside of the
PrimeProbe2 converter. Tighten the glands by hand, and then
give them an extra

½-turn using a crescent wrench.

7.

Replace the back cover onto the PrimeProbe2 converter.

After successful installing the sensor and connecting the converter to the
PrimeProbe2 sensor, configure the converter. Program a PrimeProbe2
with an integrated converter using the PrimeWorks software on a
PC/laptop. A PrimeProbe2 with a remote converter can be programmed
using either the PrimeWorks software or the 3-key membrane keypad on
the front of the converter. Instructions on programming the PrimeProbe2
using PrimeWorks are at the end of this section.