PrimeProbe2 User Manual

40

Enter the Internal Pipe Diameter and select the Sensor
Insertion Point
. ADS recommends centerline (Centre Line).

ADS recommends selecting the Energy Saving checkbox.
Leaving it unselected will cause the rapid depletion of the
battery. This option is available only when Permanent is
selected on the first screen.

6.

Click Next to go to the last wizard screen.

7.

Enter the maximum flow value either by selecting a flow rate
(Manual) or specifying a velocity (Automatic). This scales the
output pulses deriving an FCAL value that will be programmed
into the attached logger. The maximum output from the
PrimeProbe2 is 32Hz. If the maximum flow for a site is unknown,
ADS recommends selecting Automatic and setting the velocity to
7 feet per second (2 m/s). Velocity is unlikely to exceed this value.

8.

Click Program to display the FCAL value. This is the value that
will be programmed into the connected logger. When using an
ADS PrimeLog, make sure the logger is programmed with the
sensor type PrimeProbe2 and this FCAL value.