Bio-Rad Gel Doc™ XR+ System User Manual

Page 23

Advertising
background image

Page 23 of 42


1. Make sure that all lights in the Universal Hood II are off.
2. Make sure that the Lens Cap is on the lens. This is critical!
3. Power ON the Camera and the Universal Hood II.
4. Wait for 30 minutes for the system to warm up and stabilize
5. Double click on Quantity One

 software icon on the desktop. From “File Menu” select

Gel Doc

 XR OR ChemiDoc

XRS.


TDS Quantity One

Window


Select Gel Doc

XR OR ChemiDoc XRS from

File Menu


5. The following window will open with a notification that the system will be acquiring Dark

Reference images.






Advertising