Appendix E: Technical Specifications

Model

Gel Doc

XR

ChemiDoc

XRS

Application

Chemiluminescence

No Yes

Fluorescence

Yes Yes

Chemifluorescence

No No

Colorimetric/Densitometry

Yes Yes

Gel Documentation

Yes Yes

Isotopic Imaging

No No

Hardware Specifications

Maximum Sample Size

28 x 36 cm

28 x 36 cm

Maximum Image Area

25 x 26 cm

25 x 26 cm

Excitation Source

Trans UV 302, 254,

365 nm & White,

Epi-White light

Trans UV 302, 254,

365 nm & White,

Epi-White light

Illumination modes

3

Trans UV & White,

Epi-White

3

Trans UV & White,

Epi-White

Detector

CCD

Super Cooled

CCD

CCD Resolution (H x V)

1360(H) x 1024 (V)

1300 (H) x 1030 (V)

Pixel Size (in Microns)

4.65(H) X 4.65 (V)

6.7 (H) X 6.7 (V)

Cooling System

No Peltier

cooled

Camera Cooling Temp.

NA

-45

°

C

Filter Holder

3 positions

2-Emission Filter

1-Chemi

3 positions

2-Emission Filter

1-Chemi

Emission Filters

1 Included (amber),

4 optional

1 Included (amber),

4 optional

Dynamic Range

3.0 Orders

> 3.0 Orders

Pixel Density (gray levels)

12 bit

(4096 gray levels)

12 bit

(4096 gray levels)

Dynamic Flat Fielding

No Yes

Instrument Size (cm)

36(L) x 60(W) x 96(H)

36(L) x 61(W) x 96(H)

Instrument Weight (kg)

32 kg

32 kg

Software Specifications

Application Driven

Yes Yes

Windows XP

Yes Yes

Windows 2000

Yes Yes

Mac OS 10

Yes Yes

Image File Size (MB)

Approx. 2.5 MB

Approx. 2.5 MB

Operating Ranges

Operating Voltage

100/115/230 VAC

Nominal

100/115/230 VAC

Nominal

Operating Temperature

10-28°C (21°C

recommended)

10-28°C (21°C

recommended)

Operating Humidity

< 70% non-
condensing

< 70% non-
condensing