Bio-Rad Gel Doc™ XR+ System User Manual

Page 34

Advertising
background image

Page 34 of 42


Lamp replacement Starter replacement

8.3.2 Starter replacement (P/N 100-1370)

Turn off the power.

Disconnect the power cord from the Universal Hood II.

Remove the four screws located on the left/right sides of the drawer.

Remove the cover with the UV filter by sliding it forward.

Place it down on its backside on a non-abrasive surface.

Caution: Do not put the UV filter directly on the bench.

To remove the starter, rotate it counter-clockwise and pull out.

To replace the starter insert it into the holder and rotate clockwise.

Reassemble the cover and retighten the screws on both sides.

Advertising
This manual is related to the following products: