Creating a hybridization only program (hyb only) – Iris Sample Processing StatSpin® ThermoBrite Denaturation and Hybridization User Manual

Page 15

Advertising
background image

11

PGM 01 name
Enter: Temp Time
Denat: °C :00
Hyb: °C 00:00

(Reference Character Selection for additional information in regards to working with the keypad.)

Cursor will advance to

“Denat Temp”. With numeric keypad enter a two-digit temperature value in

degrees Celsius (50-99ºC).

Cursor advances to

“Denat Time”. With numeric keypad enter a two-digit time value in minutes (0 – 30)

Cursor advances to

“Hyb Temp”. With numeric keypad enter a two-digit temperature value in degrees

Celsius (30-70ºC). For room temperature hybridization enter the two-digit value 00.

Cursor advances to

“Hyb Time”. With numeric keypad enter a two-digit time value in hours (0 – 99)

followed by a two-digit value in minutes (0-59)

Display will now show entered program values. Cursor highlights

“Enter to Accept” line.

Denat: 82°C :05
Hyb: 45°C 20:00
ENTER to Accept
STOP to Abort

Press

“Enter” button to accept the program values; or press “Backspace” button to return to previous

screen to modify program values; or press the

“Stop” button to abort.

Creating a Hybridization Only Program (Hyb Only)

From the Main Screen, use the arrow keys to move cursor to

“Create a PGM” and press “Enter” button

to accept.

ThermoBrite will advance to the next available program number.

PGM 02 name

Enter: Temp Time
Hyb: °C 00:00

(Reference Character Selection for additional information in regards to working with the keypad.)

Cursor advances to “Hyb Temp”. With numeric keypad enter a two-digit temperature value in degrees
Celsius (30-70ºC). The instrument allows a temperature of 30

C or ambient temp + 5

C (whichever is

higher) for the lowest hybridization temperature. For room temperature hybridization enter the two-digit
value 00.

Cursor advances to

“Hyb Time”. With numeric keypad enter a two-digit time value in hours (0 – 99)

followed by a two-digit value in minutes (0-59)

Display will now show entered program values. Cursor highlights

“Enter to Accept” line.

PGM 02 EBV
Hyb: 55°C 01:30
ENTER to Accept
STOP to Abort

Advertising