ITC Hemochron Response Whole Blood Coagulation System User Manual

Page 17

Advertising
background image

17

Suppressing Results Display During a QC Test
Display of the clotting time during a QC test (on the screen, printed result, and in the database) can be
suppressed.

1.

Display the second page of the Supervisor menu.

2.

Press 6 to display the QC Lockout menu. Press MENU or 0 to display the second page.

3.

Press 1 to hide the results. Y is displayed.

Note: Press the numeral key again to select the alternate choice.
Note: If QC Hide is enabled, a QC test result will be displayed as Pass/Fail without a clotting
time. The actual test record with all results is stored and can be downloaded or accessed by
turning off QC Hide.

Defining a User Note
Up to nine customized notes of up to 16 characters in length can be defined. Up to two of these notes can be
selected and appended to a test record by the operator when the test is run.

Note: HRDM V. 3.0 or higher software can be used to enter notes on a PC and transfer them to the
instrument.

1.

Display the second page of the Supervisor menu.

2.

Press 7. In response to the prompt, enter the number of the note (1 to 9) to be created or
changed.

3.

The User Note screen is displayed, with the cursor positioned on the first character of the note
and the selection block positioned at the space:

4.

Press 8 to move the selection block right one character, press 7 to move the selection block left
one character, press 9 to move the selection block up one line, or press 0 to move the selection
block down one line.

5.

When the selection block is on the correct character, press YES to enter the selected character
and move the cursor to the position for the next character.

6.

Repeat Steps 4 and 5 for each character to be entered in the note.

Note: The note can be edited after characters have been entered. Press NO to move the cursor
one space to the right; press BACKSPACE to move the cursor one space to the left. When the
cursor is at the desired position, press 2 to overwrite a character with a space, press 3 to
insert a space before the selected character, or press 4 to delete the character at the selected
position. Press 1 to clear the note.

7.

When the note is completed, press CANCEL to save the note and exit the screen.

Cursor

Selection Block

Printed From ITC Intranet

Advertising