Prewarming
The test wells can be prewarmed to 37 °C ±1.0 °C on command. On completion, 3 short beeps are emitted.

Note: Refer to assay package inserts for prewarming requirements.

Automatic Shutdown
When operating from the battery, the system shuts down automatically after 15 minutes of inactivity. This
15-minute interval cannot be changed. When operating from the AC/DC Power Module, the system
automatically shuts down after an interval of inactivity defined by the supervisor.

Note: The factory default setting is 60 minutes.

All stored data is retained after an automatic shutdown.

Test Termination
A test is terminated if clot formation is not detected within 1,500 seconds after starting the test. A FAULT
>1500 message is then displayed and stored in the database, indicating the test result is outside of the
specified range.

Note: Results that are greater than the specified time are beyond the sensitivity range of the test.

They should be repeated immediately and, if confirmed, reported as greater than the maximum time.

A test is automatically terminated if after pressing START a test tube is not inserted into that well within 60
seconds or if a stable magnet is not detected in that well within 75 seconds.

Keypad

The use of each of the keys is summarized below:

Key Purpose
START 1, START 2, Switch power ON or OFF. Initiate a test when blood is added to a test tube.

MENU 1, MENU 2

Display the first (press once) or second (press twice) page of the main
menu for Well 1 or Well 2, respectively. Display following page(s) of other
screens.

CANCEL

Cancel an operation or return to the previous selection.

PRINT

Print results on an external and/or internal printer.

PAPER FEED

Advance the printer paper one line.

BACKSPACE

(Before entering Yes or No) Undo the previous keypad entry.

0 through 9

Enter PID, OID, PIN (optional), ESV serial number, and QC ranges. Select a
menu option.

YES (ENTER)

Save the response to a prompt or the entered ID or PIN.

NO

Reject the response to a prompt.

,

Left/Right Cursor Positioning.

,

Page Up/Page Down.

Printer Operations

Well Status Indicator LEDs

Number Pad/Option Selector

START

Menu Selections

Operator Answer

Keys

Display Panel

Power Indicators