5

SPECIFICATIONS
Specifications for the HEMOCHRON Response Whole Blood Coagulation System are listed below.

Dimensions and Weight

Depth

19 cm (7.5 in)

Width

27 cm (10.5 in)

Height

22 cm (8.7 in)

Weight

2.90 kg (6.4 lbs)

Operation

Test Wells

2

Timing Range

22 seconds to 1500 seconds

Incubation Temperature

37 °C ±1.0 °C

Incubation Warm-Up Time

30 seconds to 90 seconds

Full-Charge Operating Time

8 hours (minimum)

Battery Life

500 recharges

Throughput (Full Charge)

49 test cycles (at 150 sec per test)
17 test cycles (> 500 sec per test)

AC/DC Power Module

Input Power

90 to 264 VAC, 50/60 Hz, 1.2 Amps maximum

Output Power

+12 Volts DC, 3.5 Amps maximum (42 Watts,
144 BTU/hr)

Environmental

Ambient Temperature

15 to 30 °C

Note: For more technical information, refer to the HEMOCHRON Response Whole Blood Coagulation
System Service Manual.

GETTING STARTED

Unpacking and Inspection
Before unpacking the system, determine the area where the system will be located. You will need a level and
flat area that is approximately 30 cm (12 in) wide, 30 cm (12 in) deep, and 30 cm (12 in) high.

To Unpack the Instrument:

1.

Unpack the carton.

2.

Inspect each component for damage when unpacking. If damage is observed, contact your
shipper or service representative immediately.

3.

Place the instrument where it is to be located.

4.

Remove protective packaging.

5.

Examine the packaging material to be sure that the power supply, connecting cables or other
components have been removed. The materials that are provided are listed on the following page.

Note: Do not discard the packing material. It should be kept for shipping the instrument to

ITC, if repair is necessary.