Attention label – ITC Hemochron Response Whole Blood Coagulation System User Manual

Page 4

Advertising
background image

4

After a clot forms, the instrument beeps and the clotting time is displayed on the display panel. The result is
also stored in the system database with the date and time the test was performed, and the assay type. If
entered, the PID and OID are stored with the test result.

Features
The system has a number of performance and convenience features:

The system is portable for bedside use

A multi-test menu is resident on the system

Fresh whole blood or citrated whole blood can be used with the appropriate test tubes

A sample size of up to 2 mL of whole blood is required

Test name and expiration date are automatically read when ITC barcoded test tubes are used

Successful or errant results are automatically stamped with date and time

Results are available in minutes

Results are displayed appropriately as whole blood or plasma equivalent or INR (PT assay only)

Results of 600 patient tests and 300 QC tests can be stored for each well, with optional entry of
PID, OID, and user notes

Dose-response calculations are performed with the RxDx

®

module (if activated)

504 operator identification codes can be stored with OID/PIN and permissions

Operator lockout can be configured by OID, valid OID, or PIN, using HRDM V3.0 or higher
software or the keypad

QC lockout can be configured at one or two levels by time interval

Stored results can be reviewed by test type, PID, OID, or date

Stored results can be downloaded to a personal computer

System self checks are automatically performed

An ESV tube is available to check test well operation and detector electronics

A Temperature Verification Tube (TVT) can be additionally used to check test well temperature

The display is illuminated for viewing in low light

The display can indicate the percentage of battery power remaining numerically or graphically

The user is alerted when the battery is low

The system includes an on-board printer

Two external serial ports and a Centronics parallel port are provided

Patient/QC test reports can be created using a personal computer and ITC data management
software programs

ATTENTION LABEL

An attention label on the rear of the HEMOCHRON Response instrument alerts users to accompanying
documentation:

Before using the HEMOCHRON Response instrument, it is essential that the contents of this Operator’s
Manual
are read and understood by the operator.

Handle and open the container with care.

Printed From ITC Intranet