Rotor preparation, Caution – Scilogex D2012 Micro-Centrifuge User Manual

Page 10

Advertising
background image

USER MANUAL

www.scilogex.com

Page 10

7. Rotor Preparation

7.1 Prepare the samples

7.2 Inject the samples into tubes

● Do not overload samples into the centrifuge which will cause leaking.

●Do not exceed the actual capacity allowed in the user manual.

7.3 Keep the tubes balance

●Although the centrifuge can accept sample balancing by eye, we recommend that you keep this centrifuge in a well-
balanced condition to extend its life expectancy.
●Never intentionally run the centrifuge under unbalanced condition even though the allowable imbalance is not
exceeded.

CAUTION

ʽ

7.4 Inspect the rotor

Check the rotor for corrosion or scratches before using.

● Any abnormality such as corrosion or scratches are found, stop using the rotor and contact our service center.
● Only manufacturer’s rotors must be used with the unit.

● Make sure the rotor lid is securely fixed on the rotor, as well as the rotor and shaft are tightened. Otherwise, the
rotor may be moved off while rotating and cause damages of the centrifuge and rotor.

7.5 Symmetrically load centrifuge tubes to rotor

CAUTION

ʽ

Advertising