3 start the operation 8.1.4 end the operation, 2 rcf operation – Scilogex D2012 Micro-Centrifuge User Manual

Page 13

Advertising
background image

D2012 High Speed Mini Micro-Centrifuge

www.scilogex.com

Page 13

(2) Set the time

● Press button , time value flashes in the time setting mode.
● Rotate the program button to set running time from 30 seconds to 99 minutes.
● When the time displays HD, this is a continuous running mode.

Select

(1) Press button to start running
●Timer will operate once the speed setting value is reached, the screen displays the remaining run time.
(2) View and modify the operation programs
●Pressing button returns the display to the program interface and displays setting programs. Press the select
button to the desired program. When flashing, rotate parameter button to modify values. Release the button
after 5 seconds, and the centrifuge will return to normal operation mode and run according to the new value.
●If the set time value has been modified, the operation time is not affected and will continue.
(3) Warning display
●If an error occurs during the operation, the centrifuge will brake to stop automatically, and display the error code
on the time/display area. The error code can be checked in the table 10-1, and corrective actions can be applied
accordingly.

8.1.3 Start the operation

8.1.4 End the operation

(1) The centrifuge will brake when it reaches the setting time or button is pressed.

●When the rotor stops rotating, the centrifuge will start beeping to alert the operation has finished.

(2) Open the door

●The door can be released automatically when the operation has stopped.
●With the door closed, you are able to press the button to open it.
●After ending the operation, the program will store the setting parameters of this operation, and will recall these
parameters when restarting the program.

(3) Open the door and take out the rotor and samples.

8.2 RCF operation

START

STOP

START

STOP

Select

Select

Select

Door

(1) Turn on the power switch
(2) Set a RCF (Relative Centrifugal Force) value

●Press the select button and choose speed unit ×g, the speed symbol will flash into RCF value input status.
●If no button is pressed after the speed value has flashed for 5 seconds, the input mode will be shut down.
●Rotate program button to input a RCF value, RCF increment is100×g.

(3) Set operating conditions

The other operation, please refer to the section 8.1.