● Read the user manual thoroughly to properly load and use rotor.
● Do not exceed the allowable maximum speed of rotor atube and adapters etc.. Ensure the allowable
maximum speed of adapters is lower than the rotor’s maximum speed.

11. Instructions for the Rotor and Tubes

11.1 Rotor instructions

11.1.1 Rotor structure

Figure11-1 The rotor structure

11.1.2 Available rotors and adapters

The rotor can be used as follows:

Table 11.1 Rotors and adapters

Rotor type

ID

code

Tubes

Adapters

Maximum

speed

(rpm)

Maximum RCF

(×g)

Allowable imbalance(*)

Imbalance

Volume

imbalance

A12-2

01

1.5/2.0ml

tube

15000

15100

2.0g/ tube

5mm/ tube

0.2ml

PCR tube

A02P2

15000

15100

0.5ml micro

tube

A05P2

15000

15100

CAUTION

ʽ

Locking nut

Fastener

Rotor lid

Rotor block

Shaft

Tube holes