D2012 High Speed Mini Micro-Centrifuge

www.scilogex.com

Page 7

4. Installation

This section describes the instructions that you should abide when install the centrifuge to ensure your safety and the
optimum performance. Before moving the centrifuge, the rotor must be removed.

● Improper power supply may damage centrifuge.
● Make sure the power source conforms to the required power supply before connecting.

● To avoid electrical shocks, ensure your hands are dry when touching the power cord.
● This centrifuge must be grounded properly.

4.1 Location

(1) Place the centrifuge on a firmĂflat and level surface, ensure the four feet of this centrifuge stand on the counter
firmly. Avoid installing on the slippery surface or surface prone to vibration.
(2) Ideal ambient temperature is 20ć±5ć, avoid placing the centrifuge in direct sunlight if temperature exceeds 30ć.
(3) Keep clear of the centrifuge at least 10cm on both sides and at least 30cm behind it to guarantee the cooling
efficiency.
(4) Keep away from heat or water to avoid sample temperature issues or centrifuge failures.

4.2 Connection of the power cord and grounding

A minimum 10A outlet providing a sufficient ground is required, and this must meet local safety requirements.

WARNING

ʽ

WARNING

ʽ

3. Required Operation Conditions

3.1 Basic operation Conditions

(1) Power: 110V-240V, 50Hz/60Hz, 3A.
(2) Ambient temperature: 2ć~40ć.
(3) Relative humidity: ≤80%.
(4) No vibration and airflow around.
(5) No electric dust, explosive and corrosive gases around.

3.2 Transport and storage conditions

(1) Storage temperature: -40ć~55ć.
(2) Relative humidity: ≤93%.