11.1.3 Notice

The centrifuge rotor can separate samples which density lower than 2.0 g/ml. If the separated samples density is over
2.0 g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm) = Maximum speed × (2.0 (g/ml) /Sample density (g/ml) )1/2

11.1.4 Autoclaving

The rotor is manufactured in high-strength aluminum alloy material and can be autoclaved: 121ć(1.0kg/cm2)20
minutes.

11.2 Tubes

11.2.1 Cleaning and sterilizing tubes

Table 11.2 Cleaning and sterilizing conditions for tubes

O:Applicable X:Inapplicable

PA PC

PP

Cleaning

Cleaning fluids

Acidic(pH5 or lower)

X

X

X

Acidic(higher than pH5 )

O

O

O

Alkaline(higher than pH9 )

O

X

O

Alkaline(pH9 or lower)

O

O

O

Neutral(pH7)

O

O

O

Warm water(up to 70ć)

O

O

O

Ultrasonic cleaning

Neutral detergent(pH7)

O

O

O

Sterilization

Autoclaving

115ć(0.7kg/cm2)30minutes

O

O

O

121ć(1.0kg/cm2)20 minutes

X

O

O

126ć(1.4kg/cm2)15 minutes

X

X

X

Boiling

15 to 30 minutes

O

O

O

Ultraviolet sterilization

200-300nm

X

X

X

Gas sterilization

Ethylene oxide

O

X

O

Formaldehyde

O

O

O

Materials

Conditions

PA:Polyallomer; PC:Polycarbonate; PP:Polypropylene

● The lid of the rotor is made of plastic, can not be high-pressure sterilization, only ordinary sterilization can be used.

CAUTION

ʽ

11.2.2 Cleaning PC tubes

PC material is low in chemical resistance against alkaline solutions. Avoid using neutral detergents with pH
higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted according to the the
manufacturer’s instruction. Use detergent with its pH between 7 and 9.