ETS-Lindgren HI-6153 Electric Field Probe User Manual

Page 103

Advertising
background image


Operating Protocols

|

103

Probe
Command

Description

Probe Response

BP

Read probe converter
voltage in hexadecimal
format

:B64N

N=safe operating level

F=fail level

Voltage reported as 0–64;
64 corresponds to 100%

The

BP command is provided for backward compatibility and should

not be used to monitor the converter voltage, which always responds
with :B64N<CR>.

Probe
Command

Description

Probe Response

D3

Read probe data

:Dx.xxxyy.yyzzz.zB<CR>

xxxx, yyyy, zzzz=

4-digit axis values with
floating decimal point

B=battery flag, N or F

D5

Read probe data

:Dx.xxxyy.yyzzz.zcccc.B<CR>

xxxx, yyyy, zzzz=

4-digit axis values with
floating decimal point

cccc=composite field value

with floating decimal point

B=battery flag, N or F