4.0

Calibration

The

Tracer AV

can be calibrated using the

Zero

and

Units

keys, EDP commands, or the

Revolution

®

configuration

utility. Each method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights
using hooks or other additional items.

Figure 4-1. Calibration (CALIBR) Menu

You can also use the default calibration feature, which utilizes a more user-friendly procedure; however, not all
options are available in default calibration.

Figure 4-2. Default Calibration (DEFCAL) Menu

4.1

Default Calibration

If you use a 500lb or 300lb scale, you can select it
from the

DEFCAL

menu. This automatically uses the

settings shown in Table 4-1. for 500lbs or Table 4-2.
for 300lbs.

To calibrate the

Tracer AV

using the

Zero

and

Units

keys, do the following:

1. Place the

Tracer AV

in setup mode

To enter setup mode, remove the setup switch
access screw on the

Tracer AV

(See Figure 2-3

4). Once the screw is removed, use a small
screwdriver to press the setup switch. The
display changes to show the word

DEFCAL

.

2. Remove all weight from the scale platform. If

your test weights require hooks or other
items, place them on the scale.

3. Press the

Zero

key to go to

SELECT

.

4. Press the

Units

key to go to zero calibration

(

WZERO

).

5. Press the

Zero

key to calibrate zero. The

indicator displays

*CAL*

while calibration is in

progress. When complete, the A/D count for
the zero calibration is displayed. DO NOT
adjust this value after WSPAN has been set!
Press

Zero

again to save the zero calibration

value and go to the next prompt (

WVAL

).

XXXXXXX

XXXXXXX

WZERO

WV AL

WSPAN

SETPNT

PROGRM

SERIAL

CALIBR

CONFIG FORMAT

REZERO

VERS

Zero

DEFCAL

XXXXXXX

XXXXXXX

SETPNT

PROGRM

SERIAL

CALIBR

CONFIG FORMAT

VERS

XXXXXXX

DEFCAL

SELECT

WV AL

WSPAN

WZERO

500lb

300lb

Primary

500lb x 1lb

Secondary

250kg x .5kg

SP1 Trip

50lb

SP2

Off

Digital Filter

8,8,8

Zero Tracking

3D

Table 4-1. 500lb scale settings

Primary

300lb x 1lb

Secondary

150kg x .5kg

SP1 Trip

50lb

SP2

Off

Digital Filter

8,8,8

Zero Tracking

3D

Table 4-2. 300lb scale settings